Creating a Method
Creating a Method for the GC
52
ISQ User Guide
Thermo Scientific
remain at that temperature after the analytical run is complete. You can also set the
amount of pressure used to push the carrier gas through the column.
3. If you have a split / splitless inlet (SSL) injector, click the
Left SSL
or
Right SSL
tab to
configure the injector port settings. The inlet should be turned on and set to a
temperature that is at least 10 °C higher than the boiling point of your least volatile
analyte. The material should be injected into the port to vaporize and move into the GC
column quickly. Higher temperatures can lead to thermal decomposition of some
analytes, so you will have to optimize the injector temperature for your analysis. The SSL
temperature can be set between 50 and 400 °C (a typical value would be 225 °C).
Figure 39.
Setting the TRACE GC Ultra SSL Injector Parameters
a. Useful for diluting high concentrations of sample, the split flow is the amount of gas
that is swept through the injector to the exhaust port. Higher values will give more
dilution. The split flow will reduce the amount of contamination that builds up in
your system. The split flow ratio is the ratio of the split flow to the carrier gas flow. It
is effectively the dilution ratio of the sample. This setting is typically turned on and
set to a flow of 50 mL/min. However, more carrier gas will be used, so for your
analysis, lower split flows may be more acceptable.
CAUTION - INSTRUMENT DAMAGE.
Be sure not to overheat the GC column or it may
contaminate the ISQ instrument.