Changing the Column
Connecting the Column to the Transfer Line
16
ISQ User Guide
Thermo Scientific
Figure 11.
Positioning the Column in the Transfer Line
21. Tighten the nut 1 flat forward—back to where it is secure enough in the transfer line that
you cannot move it.
22. Tighten the nut 1 additional quarter turn.
23. Remove the cut septum.
24. Close the front door of the GC.
25. Condition the graphite Vespel ferrule:
a. Raise the oven temperature to the maximum temperature you will operate the GC
column.
b. Wait 10 minutes.
c.
Lower the oven temperature to 40 °C (104 °F) and allow it to cool before continuing.
d. Retighten the transfer line nut.
26. Restore working conditions.
a. Raise the oven temperature to the initial temperature that you will use.
b. Turn on vacuum compensation on the GC.
c.
Twist the vent valve clockwise to close the valve. Be sure not to pinch the o-ring.
Note
If you are using a SilTite ferrule, follow the instructions that come with SilTite
ferrules. If you are using a graphite Vespel ferrule, they require conditioning to ensure
a leak-tight seal. See the
ISQ Spare Parts Guide
for information about ordering these
ferrules.
Transfer Line
Column
Septum
WARNING BURN HAZARD:
The oven may be hot. Allow it to cool to
room temperature before opening it. The injector will still be hot, so do not touch
it.