2
Changing the Column
Connecting the Column to the Transfer Line
14
ISQ User Guide
Thermo Scientific
8. Use a scoring wafer to score and break the column. Use a magnifying glass to check for an
even, flat cut. Repeat if necessary.
9. Use a 5/16 in. wrench to hold the column measuring tool steady.
Figure 6.
Column Measuring Tool
10. While holding the column measuring tool steady, tighten the transfer line nut with a
1/4 in. wrench until the column just stops moving in the ferrule.
11. Turn the transfer line nut 1 flat backward so the column is able to move in the ferrule
with slight resistance.
12. Line up the outlet of the column with the arrows on the end of the column measuring
tool.
Figure 7.
Lining Up the Column in the Column Measuring Tool
13. Place a septum with a notch cut into it behind the transfer line nut. The septum marks
the place on the column where it should exit the nut.
Figure 8.
Positioning the Septum
14. Pull the column back from the transfer line nut. Do not move the septum from its
position on the column.
Column Outlet
Transfer Line
Nut
Septum
Column
Measuring
Tool