Configuring the Autosampler
a. Configure the Sampling group:
•
Sample Volume
—Use this field to enter the sample volume to be injected into
the GC. Typical values are between 0.5 and 5 μL.
•
Plunger Strokes
—Use this pull-down menu to select the number of plunger
strokes to use when drawing up the sample. Air bubbles in the syringe change the
amount of sample injected, which can cause signal variation in different runs. To
prevent this from occurring, increase the number of plunger strokes to reduce the
chance of an air pocket in your syringe. Typical values are between 3 and 10.
•
Viscous Sample
—Use this field to select Yes if your sample is viscous or No if
your sample is non-viscous. With a viscous sample, the syringe is filled more
slowly than if it was non-viscous. Since the amount of time saved is so small, it
may be easier to just set this option to Yes.
•
Sampling Depth in Vial
—Use this pull-down menu to select the depth (Center
or Bottom) at which the tip of the syringe needle will be placed in the sample vial
when it is being filled. The bottom is the default and most commonly used
setting.