2
Changing the Column
Connecting the Column to the Transfer Line
18
ISQ User Guide
Thermo Scientific
Figure 12.
Transfer Line Nut and Graphite Vespel Ferrule Orientation
8. Insert the column into the column measuring tool (see
), which is in the ISQ
Toolkit , so that it is even with the lines at the end of the column.
proper positioning of the column in the tool for accurate measuring.
9. Use a scoring wafer to score and break the column. Use a magnifying glass to check for an
even, flat cut. Repeat if necessary.
10. Use a 5/16 in. wrench to hold the column measuring tool steady.
Figure 13.
Column Measuring Tool
11. While holding the column measuring tool steady, tighten the spring loaded transfer line
nut with a 1/4” wrench until the column just stops moving in the ferrule.
12. Turn the spring loaded transfer line nut 1 flat backward so the column is able to move in
the ferrule with slight resistance.
13. Line up the outlet of the column with the arrows on the end of the column measuring
tool.
Spring Loaded Transfer
Line Nut
Ferrule
Flat on the
Transferline Nut