4
Creating a Method
Creating a Method for the GC
Thermo Scientific
ISQ User Guide
45
to use this feature, set the GC oven temperature, as well as the amount of time to remain
at that temperature after the analytical run is complete. You can also set the amount of
pressure used to push the carrier gas through the column.
3. If you have a split / splitless inlet (SSL) injector, click the
S/SL-Front
or
S/SL-Back
tab to
configure the injector port settings. The inlet should be turned on and set to a
temperature that is at least 10 °C higher than the boiling point of your least volatile
analyte. The material should be injected into the port to vaporize and move into the GC
column quickly. Higher temperatures can lead to thermal decomposition of some
analytes, so you will have to optimize the injector temperature for your analysis. The SSL
temperature can be set up to 400 °C (a typical value would be 225 °C).
Figure 33.
Locating the Settings for the SSL Injector on the TRACE 1300 or TRACE 1310
a. Useful for diluting high concentrations of sample, the split flow is the amount of gas
that is swept through the injector to the exhaust port. Higher values will give more
dilution. The split flow will reduce the amount of contamination that builds up in
your system. The split flow ratio is the ratio of the split flow to the carrier gas flow. It
is effectively the dilution ratio of the sample. This setting is typically turned on and
set to a flow of 50 mL/min. However, more carrier gas will be used, so for your
analysis, lower split flows may be more acceptable. If you set the split ratio, the
software will calculate the correct split flow. The reverse is true also.
Tip
We recommend turning on the septum purge, which means additional carrier gas will
go through the injector. The default purge flow value is 5 mL/min. This reduces the
buildup of contaminants in the injector, on the column, and in the ISQ instrument. If
you perform a splitless injection, even if the split flow is set, the split flow will be turned
off for the splitless time. The septum purge will be turned off for the stop purge time.
After these times, the split flow and septum purge will be reactivated.