Preface
Hydrogen Safety Precautions
Thermo Scientific
ISQ User Guide
xv
Using Hydrogen with ISQ
To use hydrogen with the ISQ instrument, you must always shut off the GC carrier gas before
venting or turning off the ISQ instrument. There are three hydrogen safety screws on the ISQ
instrument that
must
be in place. These are attached to your instrument at the factory.
Make sure all the covers and panels of the ISQ instrument are firmly attached before powering
on the ISQ instrument. If you vented the system, make sure the vent valve is tightly closed
before powering on the system. Make sure all fittings, ferrules, and o-rings are sealed prior to
powering on the system.
Figure 1.
Hydrogen Safety Screws on the ISQ Instrument
Right Top Cover
Screw
Front Panel
Screw
Left Top
Cover Screw