2
Changing the Column
Connecting the Column to the Transfer Line
22
ISQ User Guide
Thermo Scientific
28. Power on the ISQ mass spectrometer.
29. Once the ISQ instrument is pumped own and able to scan, view air water spectra and
look for evidence of leaks with a large
m/z
28 signal. If you observe a leak, stop scanning
and gently tighten the nut in small increments until no leaks appear when scanning.
WARNING - FIRE HAZARD:
If you are using hydrogen, do NOT reach over the top of the
instrument to power it on. Instead, reach around the right side or go to the back of the
instrument and flip up the power switch.