Confirming Your GC/MS System is Working
Use the information in this chapter to confirm that your GC/MS system has power, the
carrier gas rate is correct, the gas tank has enough pressure, the system has reached vacuum
and temperature, and is leak-free.
Checking Power to the System
To confirm that the ISQ system is powered on, make sure the Power light on the front panel is
solid green. If it is not lit, the ISQ system is not powered on. To power it on, reach around the
right side of the instrument and pull up the power switch on the back. If the instrument still
doesn’t power on, check the electrical connections and wall outlet.
IMPORTANT
You need to change the GC column before setting up a method. See
Chapter 2, “Changing the Column,”
for instruction on changing the column.
Contents
•
•
Verifying the Carrier Gas Flow Rate
•
Checking Your Carrier Gas Tank Pressure
•
Checking the Vacuum and Temperature
Note
Many nitrile and latex gloves not certified for clean room use contain silicone mold
releasing agents that will contaminate the instrument. For this reason, clean room gloves
are strongly recommended when you work with the ISQ. We recommend Cardinal
Health CP100 Nitrile Cleanroom Gloves. See the
ISQ Spare Parts Guide
for ordering
information.