7
Optimizing Your Method
Modifying an Automatic Tune
96
ISQ User Guide
Thermo Scientific
–
50
—Select this option to measure the width at 50% of the peak height.
•
Threshold
—Use this field to enter the change in intensity that has to occur when the
tune to select a new voltage. For example, if you set this field to 1.1, the tune will not
select a new voltage for that component, unless the intensity is 110% of the old
intensity. If you set this field to 1, anytime the new voltage has an intensity larger
than the old intensity, the tune will select the new intensity.
•
On Error
—Use this pull-down menu to select how to handle an error in the tune.
–
Fail
—Select this option to stop the tune when an error occurs or if no tune
points are found meeting the tune criteria.
–
Continue
—Select this option to allow the tune to continue on to the next device
when an error occurs.
6. Click the
Resolution
tab to configure the resolution of your tune. You can tune the
resolution during a lens tune or you can tune the resolution by itself.
•
Tune Resolution
—Select this option to tune the resolution by itself. This resolution
will tune the system at 100 and 1000 amu/s scan rates. You may also tune at a higher
scan rate.
Note
If a value less than 1 is chosen for ramped devices, the tune will step back in energy
until the lower intensity is detected. For example, if 0.8 is chosen, the device value lower
in energy than the device value with the maximum response which gives an intensity
closest to 80% of the maximum intensity will be chosen.
Note
Click the
Advanced Settings
check box to access the
Resolution
tab.