Electrical data
Power rating
W
1800
3100
3100
1400
3360
3360
Maximum current
A
7.9
13.5
13.5
6.1
14.0
14.0
Earthing system (e.g.
1/N/PE)
1/N/PE
1/N/PE
1/N/PE
1/N/PE
1/N/PE
1/N/PE
Power line frequency
Hz
50
50
Power line voltage
+/- 10 %
V
230
230
230
230
IP protection system
IP 20
IP 20
Protection class
I
I
Overvoltage category
II
II
Device fusing, building
side
A
16
16
Device fusing, on PCB
A
2 x 16
2 x 16
Environmental conditions
Min. ambient temperature
°C/°F
18/65
18/65
Max. ambient temperature
°C/°F
32/90
32/90
Max. humidity in service,
non condensing
% r.F./
% r.H.
80, non condensing
80, non condensing
Min. storage temperature
°C/°F
20/68
20/68
Max. storage temperature
°C/°F
60/140
60/140
Max. humidity in storage,
non condensing
% r.F./
% r.H.
90, non condensing
90, non condensing
Post-transport acclimation
time
h
2
2
Noise level
dB(A)
no inherent noise
45
52
52
Pollution degree
2
2
Site conditions
Maximum altitude above
sea level
m/y
ASL
2000/2187
2000/2187
Minimum side clearance
mm/in
50/2
50/2
Minimum front clearance
mm/in
590 / 23.2 690 / 27.2
814 / 32
590 / 23.2
690 / 27.2
814 / 32
Minimum back wall
clearance
mm/in
80/3.2
80/3.2
Minimum bottom
clearance
mm/in
200/8
200/8
Minimum top clearance
mm/in
300/12
300/12
Table 13-1Technical Data - OGS and OMS Series
Parameter
Unit
OGS 60
OGS 100
OGS 180
OMS 60
OMS 100
OMS 180
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...