Handling and Control
Settings
7-10 Heratherm
Thermo Scientific
Calibration
The
Settings -> Calibration
menu item enables the user to initiate a temperature
calibration routine for the built-in temperature sensorsand choose whether calibration
should be accomplished manually or automatically:
• The
Manual
option allows for entering an absolute temperature directly, as measured -
for example, using an external reference sensor.
Press
to go to the next entry (or
to go back to
the previous one).
After reaching the entry numbered
22
the display wraps
and returns to the beginning of the error log, that is, to
entry “number
E01
.”
To exit from the error log and return to normal display
mode press
twice.
The
Settings
icon in the menu bar will go out.
Table 7-10Reading the Error Log
Calibration Prerequisites
Maintain the ambient conditions within the specified limits of the oven
and make sure that the damper is fully closed before launching
calibration.
Varying ambient conditions and/or an open damper may impact the
result of the calibration routine, which may lead to misadjustment of
the controller and unreliable temperature control operation.
Table 7-11Entering the Calibration Reference Temperature Manually
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to
confirm.
Press
to switch to the
CAL(ibration)
menu item.
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...