Installation
Space Requirements
Thermo Scientific
Heratherm 3-3
Mark hot exhaust air piping with appropriate hot surface warning signs or restrict access to
such piping altogether to avoid the risk of injury through contact with hot surfaces.
Space Requirements
When installing the oven, make sure that the installation and supply connections remain freely
accessible.
The specified side clearances represent minimum distances.
Figure 3-1
Oven Dimensions and Required Clearance
Table 3-1 Heratherm Oven Dimensions
Type
A
(mm/inch
*
)
*
Dimensions in inches are rounded equivalents specified for information only. Depth of handle /display (66 mm/2.6 in) not
included in overall depth specified; height of adjustable feet (36 mm/1.4 in) not included in overall height specified.
B
(mm/inch)
C
(mm/inch)
D
(mm/inch)
OGS 60 / OMS 60
530 / 20.8
565 / 25.2
720 / 28.3
540 / 21.3
OGS 100 / OMS 100
640 / 25.2
565 / 25.2
820 / 32.3
650 / 25.6
OGS 180 / OMS 180
640 / 25.2
738/ 29.1
920 / 36.2
650 / 25.6
Table 3-2 Required Clearances
E
(mm/inch)
F
(mm/inch)
G
(mm/inch)
H
(mm)
80 / 3.2
50 / 2
300 / 12
300 / 12
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...