Thermo Scientific
Heratherm 9-1
9
Cleaning and Disinfection
Cleaning
Cleaning exterior surfaces
Remove dirt residues and depositions thoroughly using a solution of lukewarm water and
commercial detergent.
Wipe the surfaces clean using a clean cloth and clear water.
Then, wipe the surfaces dry using a clean cloth.
Wipe / Spray Disinfection
The manual wipe and spray disinfection is the following process:
• predisinfection,
• Cleaning as appropriate for the current application.
Incompatible cleaners
Some oven components are made of plastics. Solvents may attack
plastics. Strong acids or alkaline solutions may cause
embrittlement of plastics.
Moisture-sensitive components
Do not spray cleaning agent onto the control panel and the control
box at the rear of the oven. When wiping the oven clean, always
make sure that no moisture enters into these components.
Wipe the display window down with a slightly dampened cloth,
then wipe dry with a rag made of 100% microfiber.
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...