Handling and Control
Thermo Scientific
Heratherm 7-3
The table below contains brief descriptions of the menu bar icons (itemD2
in
.x
Table 7-2 Display Features
Feature
Item
Function
Display pane showing a permanent readout of the actual tem-
perature in the work space either in °C or °F (depending on the
user's preferences, see
“Toggling the Temperature Display Unit”
A flashing time entry prompt of the general format
hh:mm
(hours:minutes, both with two digits) appears in this place while
the user is setting up a timer.
Upon occurrence of an error condition, the current error code
flashes in this area. Additionally, the red alarm icon at D3
flashes.
Menu bar with iconized representations of adjustable parame-
ters. A red border is used to highlight the current menu item, as
selected using the
Menu
Left
Right
(K4). Brief descriptions of the individual menu items are
given in
Note
If a menu item cannot be selected, then the function it represents is not
part of the equipment configuration of your unit.
Alarm icon: Upon occurence of an error condition, the red alarm
icon will be illuminated. At the same time the current error code
will flash in the display pane D1. The alarm may be acknowl-
edged by pressing the
button.
Table 7-3 Menu Bar Icons
Icon
Function
Temperature Set Value
Allows for changing the temperature set value (factory-preset to
50 °C/122 °F
) within the permissible temperature range The set value can
be changed by pressing the
Left
and
Right
(item
and you can,
after confirming your changes with the
Menu/Enter
track the impact on the actual temperature in the display pane at
Instructions:
Timer
Allows for having the oven turn on and/or off upon expiry of a user-speci-
fied countdown period. When the user enables an “on timer” the oven
goes out. A rotating in the
Timer
icon and the illuminated readiness indi-
cator icon in the status display area indicates that the timer is running.
Instructions:
Settings
Invokes a submenu with the following functions:
- Read access to error log
- Calibrating the oven
- Toggling the temperature display unit between °C and °F
- Entering a configuration control code
(Instructions:
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...