Maintenance
Replacing the Power Cord
Thermo Scientific
Heratherm 10-3
No tools are required to replace the seal.
Figure 10-1
Door seal replacement (example showing OMH Series oven)
1. Pull the seal out of the guide slot.
2. Starting on the hinge side of the door, position the seam of the new seal at the location
indicated by the arrow in
3. Gently press the seal into the slot, working around the circumference of the door. In doing
so, be careful not to stretch the seal.
4. Make sure that the retaining rail taper is positioned correctly in the slot and that the seal is
flush with the door frame.
Replacing the Power Cord
When the device’s power cord is damaged, it must be replaced with an original spare part.
Using a standard power cord with a lower temperature withstand class is prohibited.
Returns for Repair
Prior to returning any materials, please contact our Customer Service Department for a
“Return Materials Authorization” number (RMA).
Material returned without an RMA number will be refused.
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...