Product Description
Safety Devices
4-6 Heratherm
Thermo Scientific
Safety Devices
The oven is equipped with the following safety features:
• a sample protection feature that safeguards the samples against destruction through
overheating in case of contoller failure;
• an overheat protection cut-out feature that shuts down the oven completely when
excessive temperatures occur in the workspace;
• dual fuses rated at 16 amperes.
Operating Environment
To ensure undisturbed operation, the ambient temperature in the operating room must be at
least 18 °C (64.4 °F). The heating system controls the temperature in the oven’s workspace
of 50 °C/122 °F up to the maximum of 250 °C/482 °F .
Sensing and Control System
The PT 100-type sensor for the control of the work space temperature and for the thermal
protection [1] is mounted in the top panel of the work space compartment.
Figure 4-5
Sensor Mounting Location - OGS Series and OMS Series
The work space temperature sensor provides the inputs to the oven’s built-in controller,
which continuously compares the measured values to the user-specified set value and adjusts
the heaters according to the result.
The unit features a thermal protection function that is factory-preprogrammed and not
adjustable. It protects the samples in the work space from overheating: Thermal protection
kicks in on a brief violation of the upper limit, based on the user-selected temperature set
value, automatically reducing the work space temperature to the user-specified set value and
allowing the heating process to continue even in case of a controller malfunction. If the
thermal protection is activated, the error message (E111) “Temperature too high” appears in
the display window and an audible alarm is sounded.
Содержание Heratherm OGS 100
Страница 1: ......
Страница 6: ...Contents 4 Heratherm Thermo Scientific ...
Страница 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Страница 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Страница 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Страница 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Страница 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Страница 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Страница 84: ...Device Log 15 2 Heratherm Thermo Scientific ...
Страница 87: ...Contact Thermo Scientific Heratherm 16 3 ...
Страница 88: ...Contact 16 4 Heratherm Thermo Scientific ...