Data Communication
USB Interface
The units come equipped with a USB interface. The USB interface complies with Standard USB 1.1 /
USB 2.0 / USB 3.0 (full speed). The USB port is operated as a virtual COM port. Therefore, the port
transmission speed can be changed within the defined baud rates (9,600, 19,200, 38,400, 57,600
baud. Data exchange is accomplished via a defined command sequence structure (frames). The
command sequences correspond with the RS 232 interface installation diagram.
Contents
•
“Structure of the Command Sequences for Data Communication”
•
“Overview of General Parameters (Addresses 0xxx)”
•
“Overview of Incubator Parameters (Addresses 2xxx)”
•
•
•
“Examples of Data Logger Codes”
•
“Program Heracell Vios 160i / 250i CR”
Note
Installing the USB port with the virtual COM port:
If the USB port is to be used for data exchange between a computer and the incubator, the USB
port is installed as virtual COM port (USB serial port) using the supplied driver.
The assigned COM port can be located in the Windows Device Manager/Ports dialog box, e.g.
USB Serial Port (COM5), and is then defined as communication port in the
Heracell Vios 160i /
250i CR
“Program Heracell Vios 160i / 250i CR”
).
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...