8
Cleaning and Disinfection
Steri-Run Sterilization Routine
8-12
Heracell Vios 160i / 250i CR
Thermo Scientific
Steri-Run Interruption due to Error
If an error occurs while the sterilization routine is run, an error message is displayed and the following
actions are initiated:
• the sterilization routine automatically changes to the cool-down phase,
• the audible alarm sounds.
1. To confirm the audible alarm:
• Press any position on the display.
• The audible alarm is switched off. The
End
key is displayed. If the sterilization routine is not
canceled after this, cool-down to the set temperature occurs.
2. Cancel Steri-Run:
• Press the
End
key.
• The error message is displayed.
• If the error message is confirmed, the system returns to the main menu.
Figure 8-7.
Interrupting / Canceling Steri-Run
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...