6
Handling and Control
Error Messages
6-50
Heracell Vios 160i / 250i CR
Thermo Scientific
*1)
This error appears on the display only; it is not entered in the error log.
0...100%
O
2
Sensor breakage
Measured value exceeds
accepted limit
Call field service
X
X
X
Actual value above
Actual value > Set value
+ 1%
*4)
Check the gas supply
source. Reduce
primary pressure to
1 bar (14.5 psi) max.
X
X
X
Actual value below
Actual value < Set value
- 1%
*4)
Check the gas supply
source. Replace gas
cylinder. Increase
primary pressure to
1 bar (14.5 psi) max.
Check feeding line.
X
X
X
Error communication
Sensor not
communicating with
the main board
Call field service
X
X
X
Failure: Gas cylinder
changeover switch
Gas cylinder
changeover switch not
communicating with
the main board
automatic
X
X
X
No gas
Both O
2
gas cylinders
are empty
Swap at least one O
2
cylinder.
X
X
X
Gas cylinder A empty
Gas cylinder A is empty Replace gas cylinder
A
Gas cylinder B empty
Gas cylinder B is empty Replace gas cylinder
B
rH
No water
Not enough water in
water reservoir.
Refill water or, if dry
operation is desired,
de-activate the water
level sensor. In case
the failure message
occurs again after
filling, troubleshoot
the device according
to Section
error persists, call
field service.
X
X
X
Fan
The LED on the
outside lights up red
Fan in the rear wall
enclosure doesn‘t work
properly
Control
loop
Failure message
Cause
Remedy
Alarm
relay
Audible
alarm
Log
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...