6
Handling and Control
User Configuration
Thermo Scientific
Heracell Vios 160i / 250i CR
6-35
Activating/De-activating the Water Level Sensor
For incubation operation with ambient humidity or if the auto-start routine is to be run dry (without
water), the water level sensor can be switched off. This prevents alarm messages for the water level
sensor by the device-integral monitoring system from being issued:
1. Press the
Menu
key.
2. Select the menu command
Options
.
• The selection dialog shown in
is displayed.
3. Select the menu command Water level sensor.
• The input dialog shown in
is displayed.
Figure 6-40.
Setting the Water Level Sensor
1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Switching the Audible Alarm on/off:
If the device-integral monitoring system detects an error:
• an audible alarm sounds in addition to the visual error message
• in addition to the visual error message and the switching of the alarm relay.
The audible alarm can be permanently disabled.
1. Press the
Menu
key.
2. Select the menu command
Options
.
WARNING
When the water level sensor is de-activated, the Steri-Run routine can be started
despite of the presence of water in the water reservoir. This is improper usage and may destroy the
fan motor.
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...