6
Handling and Control
User Configuration
Thermo Scientific
Heracell Vios 160i / 250i CR
6-37
Figure 6-42.
Switching the O
2
Control on and off
1. To toggle between two states of the O
2
control:
• Press the + key.
or
• Press the - key.
2. The value change appears in the display. The New message indicates that the selection has been
changed, but not saved yet.
3. To accept and save the setting:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Airing the Workspace
If the unit is operated with an O
2
or N
2
supply, the workspace must be vented after de-activating O
2
control.
Note
Display of O
2
value:
When O
2
control is de-activated, the actual value readout shows three dashes (- - -) in the O
2
display field.
This procedure provides protection for the O
2
sensor. If the set value is set to 21%, the O
2
control
loop is not monitored. This is true for both O
2
control ranges:
• Control range I: 1 % to 21 %
• Control range II: 5 % to 90 %
In this case, the O
2
display shows the actual value.
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...