3
Description of the Device
Supply Interface
Thermo Scientific
Heracell Vios 160i / 250i CR
3-15
The table below lists the output signals:
The measured samples are governed by the following function:
I
0
= 4m
A,
MV = measured value (°C, % or rH%)
I
max
= 20m A,
MB = measurement range
The meaning of the output signals is explained in the table below:
The four 4-20 mA output signals are broken out onto a detachable, 8-pin terminal block (3,5 mm /
0.14 inch). The mating connector is delivered as a standard accessory. The pin-out can be taken from
the following table.
Channel
No.
Output Parameter (actual
value)
Resolution
Measuring
range
Output
signal
1
Temperature
0.08 MA/°C
0...200 °C
4...20 mA
2
CO
2
concentration,
0.8 mA/%
0...20% CO
2
4...20 mA
3
O
2
concentration*
0.16 mA/%
0...100% O
2
4...20 mA
4
Not used
Parameters marked with an asterisk “*” are only output if the respective option is installed.
Current
Significance
4 mA
The measured value is smaller than or equal to the lower boundary of the
measurement range.
20 mA
The measured value is greater than or equal to the upper boundary of the
measurement range.
2 mA
The unit is fully functional, but there is no valid measured value available (for
example, the optional sensor is not installed, the unit is still in the warm-up
phase, or similar.).
0 mA
Problem
Pin No.
Channel Description
1
Channel 1: Temperature
2
Channel 1 Ground
3
Channel 2: CO
2
4
Channel 2 Ground
5
Channel 3: O
2
6
Channel 3 Ground
7
Channel 4: Not used
8
Channel 4 Ground
MW
I
0
I
I
max
I
0
–
MB
--------------------
+
=
Содержание Heracell Vios 160i CR
Страница 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 36: ...2 Installation of the Device Retrofitting Modifications 2 12 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 60: ...3 Description of the Device Electromechanical Door Lock 3 24 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Страница 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...