Thermo Scientific
GENESYS 10S Vis User Guide
91
20
Kinetics
The Kinetics application lets you measure the change in the sample absorbance as a function
of time. The local control software allows the determination of a linear rate over a particular
region, which can be defined after the data acquisition. Frequently used in enzymatic kinetics,
a factor can be multiplied by the slope of the linear rate fit to determine activity.
Computer software control lets you greatly expand the kinetics capabilities of your
instrument:
• Multicell parallel kinetics lets you monitor up to five reactions simultaneously. Extended
kinetics data acquisition exceeds the 400 data point limit of the embedded software.
• Software control streamlines the use of more sophisticated computer applications to
analyze kinetics data after collection.
Use Kinetics for:
•
•
•
•
•
•
Recalling and Recalculating Graphical Kinetics Results
•
Rescaling and Recalculating Tabular Kinetics Results
• Modifying the scale of the plot
You can work with graphical or tabular data and perform the same functions with either.
However, the location of the function keys depends on the display type.
To get started, press the
Test
button, highlight
Kinetics
and press
Enter
.
Note
The Kinetics application lets you measure only one sample at a time.
Note
The Kinetics application lets you collect up to 400 data points per run. When
setting test parameters, select the interval time and total run time accordingly.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...