23
Parameters
Thermo Scientific
GENESYS 10S Vis User Guide
121
Linearity Value
Enters a linearity value
(Kinetics only)
To help determine the linearity of the reaction during the
measurement, the instrument offers a linearity parameter. This
is the difference between the changes in the absorbance of two
measurements as shown in the following example:
Time
Abs
ΔA
Linearity
1
.1
--
---
2
.2
.1
---
3
.29
.09
P
4
.38
.09
P
5
.46
.08
P
6
.52
.06
F
Linearity is the ΔA between ΔA calculations.
P=Pass and F=Fail
Load Test
Loads the highlighted test from the Stored Tests Directory into
active memory and sets the instrument to the test parameters
(Utility)
Lock/Unlock
Used to protect stored tests from accidental deletion or
alteration; asks for a password to allow the user to lock or
unlock the file
(Utility)
Low/High Limits
Enters the lowest and highest acceptable results, outside of
which the result is flagged as “Low” or “High”
(Adv. A-%T-C, Std Curve, Abs Ratio, Abs Diff, Kinetics, 3-Pt
Net and some Bio Tests)
Math
Accesses manipulation functions of the graph
(Scanning)
Measure Blank
(as function key)
Initiates measurement of the blank
Measure Blank
(as test parameter)
Selects the frequency of zeroing the instrument
Once or Every Reading
(Kinetics)
Parameter
Description
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...