This test lets you verify that the internal printer is functional. To run the test, you will need to
have an internal printer installed. Running the internal printer test takes no more than 20
seconds after you press Stop.
Y
To run the Internal Printer test
1. From the Utility screen, verify that the internal printer is installed properly and is
selected.
If necessary, press
Utility
and then select the internal printer.
2. In the Performance Verification screen, highlight
Internal Printer Test
.
3. Press
Enter
.
4. Press
Start Test
.
You can press
Stop Test
to stop the test.
The test print routine appears on the printed results.
If the test fails, follow these guidelines:
• Make sure the internal printer is the selected as the printer device on the Utility screen.
• Make sure the internal printer is installed correctly. (Return to the main screen and press
Enter
. If the paper does not move, the printer may not be installed correctly.)
• Make sure the thermal paper is threaded with the thermal side toward the printer head
(the outside surface of the roll is the thermal surface).
• For more troubleshooting advice contact our sales or service representative in your area or
use the information at the beginning of this document to contact us.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...