22
Maintenance
Changing the Fuse
112
GENESYS 10S Vis User Guide
Thermo Scientific
Micro flowcells can be kept clean by:
• Flushing well with a solvent after use.
• Aspirating dilute acid, base, non-filming detergent or Clorox® through the cell in short
bursts.
• Storing with distilled water in the cell.
Cleaning the Windows of the Sample Compartment
Do not use acetone or abrasive materials to clean the windows of the sample compartment.
Instead, use a non-abrasive laboratory cleaning solution (such as a commercial cell cleaning
solution), distilled water or alcohol.
Use the liquid and a soft, lint-free cloth to clean the windows. Do not apply too much
pressure or the surface of the windows may be damaged. Be sure to remove all fingerprints.
Changing the Fuse
The fuse is located in the power entry module located at the center of the back panel of the
instrument.
• 120 VAC, 2.5 A, Slo Blo
• 240 VAC, 1.25 A, Slo Blo (2 required)
Y
To change the fuse
1. Turn off and unplug the instrument from the wall outlet or power strip.
2. Position the instrument so you can access the power entry module on the back of the
instrument.
3. Remove the power cord.
4. Insert a flat-blade screwdriver into the notch on the fuse cover and pry off the cover.
IMPORTANT
Do not dry cells in an oven.
IMPORTANT
The instrument fuse must be replaced with the same type and rate.
IMPORTANT
If the fuse fails repeatedly, it may indicate a serious problem with the
instrument. Contact technical support as soon as possible.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...