3
Accessories
Cell Holders and Cell Holder Accessories
18
GENESYS 10S Vis User Guide
Thermo Scientific
Cell Holder Initialization
When you press
Run Test
to start a measurement, the instrument displays the message
“Calibrating and Checking Turret, Please Wait” while it attempts to initialize the cell changer
at the Blank position.
If a single cell holder has been installed, the message “Error, Single Cell Holder found. Use
Single Cell Holder?” appears. Press
Accept Change
to continue, or install the cell changer and
press
Cancel Change
.
If the 6-Position Cell Holder has been installed, it will initialize it to the “B” position. After it
has initialized the cell changer, the instrument will display the data collection screen for the
test.
Changing Cell Holders
To:
• use long pathlength cells (cylindrical or rectangular)
Combination Systems
+
+
+
Must be installed in the B, 2 and 4 positions
Not applicable
Cell Changer System
Single Cell System
Note
If, while in this screen, you remove the cell changer and press Run Sample, the
instrument will display “Fatal Error: 8 Press Esc to return to main menu.”
Pressing
Esc
returns you to the parameter menu screen of the test. Press
Run Test
and the
instrument will now display the message, “Calibrating and Checking Turret, Please Wait”
while it attempts to initialize the cell positioner at the Blank position.
If the cell changer is still not in the sample compartment, the instrument will display
“Error, Single Cell Holder Found. Use Single Cell Holder?”
Either press
Cancel Change
and reinstall the cell changer or press
Accept Change
to
continue with using the single cell holder.
To prevent the Fatal Error whenever removing the cell changer, always return to either the
main menu or the test parameter menu of the test before removing the cell changer.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...