13
Concentration Measurements—Advanced A-%T-C
Entering a Factor
Thermo Scientific
GENESYS 10S Vis User Guide
55
4. Press
Measure Blank
.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
5. Press
Measure Standard
.
The instrument measures the absorbance of the standard and displays the absorbance and
calculated factor.
Entering a Factor
Y
To enter a factor
Highlight
Factor
. To change the factor, enter the correct factor.
To change the units, highlight
Units
and select the correct units.
Measuring Samples
Y
To measure a sample automatically (using Auto 6 or Auto 3)
1. Press
Run Test
.
2. When prompted, place the blank and samples in their cell positions and press
Enter
.
The instrument measures the blank and samples and displays the sample measurements.
3. Press
Measure Sample
to measure additional samples.
Y
To measure samples manually (using Manual 6 or Single Cell Holder)
1. Press
Run Test
.
2. Insert the blank and sample.
A 6-Position Cell Holder can hold five samples.
3. Press
Measure Blank
.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
4. Press
Measure Sample
.
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the samples manually.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...