Cell Correction
Every test setup screen provides access to Cell Correction.
Before running Cell Correction:
• Clean the inside and outside of all the cells to be matched.
• Fill the cells with distilled water (or other blank solution), and place them in the sample
compartment (see
“Selecting and Positioning Cuvettes”
). Place the blank
cuvette in Cell “B” of the cell changer.
Cell Correction
Y
To run Cell Correction
1. Load the test type or stored test.
2. If Cell Correction is not visible, highlight
More parameters
and press
Enter
.
Note
Cell Correction is not active from the Main (Basic ATC) screen.
Note
Cell Correction is active only when the 6-Position Cell Holder is set to either Auto
6 or Auto 3. The feature is not active when the cell holder is set to 1-Cell or Manual 6, nor
when the Single Cell Holder is installed.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...