Thermo Scientific CryoStar NX70 Operator Guide
108
Transportation of the Instrument
Conditions for the Transportation of the Instrument
To guarantee trouble-free function of the instrument after transportation, please note the below-
mentioned measures for the transportation preparation.
In addition, the conditions for storage and transportation must be observed during the entire
transportation - see Technical Specifications.
Disinfect the instrument before the transportation and remove all section waste and tools
out of the chamber - see Recommended Disinfection Procedure.
Shutting down the instrument for transportation:
•
Turn off the instrument.
•
Unplug the unit.
•
Remove the blade from the blade holder.
•
Remove the section waste tray, the brush shelf, the blade holder and all other accessories as well as
tools from the Cryochamber.
•
Dispose of the defrosting and Cold D liquids in accordance to the laboratory regulations and
clean the bottle. - see Emptying the Defrosting and Cold D Liquids.
•
Remove the defrosting liquid bottle.
Note
Clean and disinfect all accessories according to the respective applicable lab regulations and transport them
in dry condition.
Any transportation of the instrument requires original packaging materials.
Note
If the original packaging is no longer available, please contact your local Thermo Scientific representation.
•
Clean and disinfect the Cryochamber - see Recommended Disinfection Procedure.
•
Unscrew the hand wheel handle for transportation if necessary.
•
Keep the heated sliding window closed during transportation.
Note
If the new site of installation can be reached in less than half an hour, keep the sliding window closed so that
the Cryochamber does not heat up. This avoids the formation of condensation water inside the chamber.
Two persons are necessary for the transport of the instrument.