Thermo Scientific CryoStar NX70 Operator Guide
18
Chapter 3 - Instrument Setup
Unpacking and Repacking
Safety Precautions
The CryoStar NX70 is heavy and requires a minimum of two people to safely unpack and
manoeuvre it.
The CryoStar NX70 has a high centre of gravity and care should be taken when moving.
Do not tip the CryoStar NX70more than 30° from the upright position.
After moving the CryoStar NX70, wait at least 8 hours before switching on to allow the
refrigerant to settle - failure to do so may cause damage to the unit.
Note
The packing materials should be carefully stored for use in any future instrument shipment.
Unpacking Procedure
General Information:
•
Ensure that these Instructions are retained for future use.
•
The CryoStar NX70 is delivered on a pallet. The sides of the instrument are encased in cardboard
and cushioned with foam spacers inside the container. A wooden cover protects the top of the
instrument.
•
In unpacking be aware of the weight (200 Kg/440 lbs) of the instrument.
Safe handling of the installation requires the involvement of two people
Handling
:
•
Place the crate on a flat surface while ensuring that there are no objects underneath the pallet.
•
Remove the wooden cover form the crate but retain it for later use as a ramp.
•
Remove the top foam spacer from the crate.
•
Remove the exterior cardboard container by opening the clips.
•
Remove the remaining foam spacers and the accessories carton. Place all the accessories carefully
to one side.
•
Remove the plastic cover from the instrument.
•
Two wing-nuts are now visible at the front edges of the wooden pallet.