Thermo Scientific CryoStar NX70 Operator Guide
96
Programming Cold D
In addition to the instant use, automatic Cold D cycles can be programmed.
Follow the directions below to activate (or deactivate) the automatic Cold D feature.
•
Press the "Menu" button on the Main screen.
•
Press the "Settings" button.
•
The Settings Menu screen will be displayed.
•
Press the "Disinfection" button.
•
The Menu: Settings: Disinfection screen will appear.
•
In this menu, the user can specify the day(s) and the specific time when the Cold D process
should automatically be carried out.
To set the Day:
•
Press the “Day” button.
•
The Menu: Settings: Disinfection day selection screen will appear.