Thermo Scientific CryoStar NX70 Operator Guide
107
Adjustment of the Movable Specimen Head Jaw
The clamping mechanism can be adjusted in case the chucks cannot be clamped properly.
•
Lock both the electronic and mechanical handwheel brakes.
•
Turn the adjusting ring by using the pin, provided with the standard accessories.
•
Turn the pin counter clockwise to increase the tension of the moveable clamping jaw.
•
By turning the pin counter clockwise, a click can be heard as a pin moves over the teeth on the
ratchet wheel. After two clicks, stop turning and check the clamping tension with a specimen
chuck. If the tension is not sufficient, increase the tension by a further click and re-test the
clamping tension.
•
Repeat this procedure until the right clamping tension is reached.
•