Thermo Scientific CryoStar NX70 Operator Guide
105
Exchange the Vacutome Filter Bag
The Vacutome Filter Bag collects all section waste which has been removed with the Vacutome function.
Section waste can be hazardous! Dispose of the used Vacutome filter bag in accordance to
the laboratory regulations.
The Vacutome Filter Bag is placed under the cover on the right hand site of the Cryochamber.
•
Open the lever of the cover.
•
Remove the Vacutome hose (1) and lift
off the cover (2).
•
Clean the Vacutome hose with a flexible brush.
•
Slip off the old Vacutome filter and replace it by a new one.