Rotors and accessories
30
BIOshield
®
- Rotor
The bucket set is a permanent part of the rotor and
must not be interchanged with other rotors.
Do not run the rotor without the rotor lid closed. The
rotor lid is opened respectively closed by pushing
down and turning the lid locking knob at the same
time.
Store the BIOshield
®
Rotor with the lid
removed after cleaning. This will enable
the rotor to dry completely.
Exceeding the maximum number of
cycles can result in rotor fracture
leading to a destruction of the cen-
trifuge!
max. cycles
22000
31000
max. speed
5850 rpm
5350 rpm
The rotor has to be replaced upon
reaching the end of its service life
depending on the mechanical load
and on the speed respectively.