Troubleshooting
76
Error message
Symptoms
Possible causes and corrective measures
E-25
Rotor stops without de-
celeration to standstill.
Start without rotor, or improper rotor fastening.
1. Switch the instrument off and on again.
2. Open the centrifuge by pressing the “lid open”-key .,
3. Check, whether the rotor has been inserted and fastened cor-
rectly.
If the error persists, contact customer service.
E-27
Rotor stops without de-
celeration to standstill.
Instrument cannot be
operated.
Mains voltage too low.
Electronics overheated.
1. Turn instrument off and pull mains plug.
2. Check and clean ventilation slots if necessary.
3. After about 60 min you can restart the instrument.
If the error persists, contact customer service.