Rotors and accessories
35
Temperature behavior:
Rotors of carbon fiber composite material require
significantly more time to change temperature than a
metal rotor does, since this carbon fiber composite
material works like a thermal insulator.
Therefore, it is recommended to store the rotor, the
adapter and the samples already at the desired run-
ning temperature.
Maintenance and care:
Do not immerse the rotor in liquid. Liquid can get
trapped inside the cavity of the bottom plate.
In case liquid has been trapped inside, a
5 minute run at 300 rpm will remove it.
Procedure in case of damage:
In case of rotor failure wrong handling can cause
dangers by the carbon fibers of the rotor.
In case of damage allow 30 minutes for
particles to settle before accessing the
rotor chamber.
Particles have to be removed from the
surface with humid cloth or sponges.
Do not sweep or vacuum dry!
Dispose rotor pieces and particles in such a way, that
an accidental exposure to fibers is not possible.
The speed must not exceed 300
rpm, otherwise rotor damages
can occur!
Wear protective clothes!
Always handle centrifuges with a de-
stroyed carbon fiber rotor with protec-
tive glasses, respirator and suitable
protective gloves of leather.