Maintenance and care
63
These agents contain highly aggres-
sive hypochlorite’s and must not be
used with aluminum rotors!
Rotor and rotor chamber must be treated with a neu-
tral, universal disinfectant. Best suited for this pur-
pose are disinfectant sprays, ensuring that all rotor
and accessory surfaces are covered evenly.
You may disinfect the rotor and the accessories as
described in the following section. Be sure to fol-
low the pertinent safety procedures for handling in-
fectious material.
1. Pull mains plug.
2. Unscrew the rotor chuck.
3. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft.
4. Remove the centrifuge tubes and adapters, and
disinfect them or dispose of them as necessary.
5. Treat the rotor and the rotor lid according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified action times!
6. Turn the rotor head down and drain off the disin-
fectant. Thereafter thoroughly rinse rotor and lid
with water.
7. Dispose of the disinfectant according to valid regu-
lations.
8. Aluminum rotors have to be treated with anticorro-
sive protective oil subsequently.
Disinfection with bleaching lye
Please note the safety measures and
handling hints when applying these
substances!