Operation
59
Grab rotor with both hands and pull
upwards perpendicularly.
Removing the rotor
1. Open the centrifuge lid.
2. Remove the rotor lid if necessary.
3. Unscrew the clamping sleeve counterclockwise
using the socket wrench supplied with the instru-
ment until no resistance exists.
1. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft carefully. Make sure
not to tilt it.
When using an aerosol-tight bio-containment lid, in
case of contamination, you may remove the respec-
tive rotor from the drive shaft without opening the lid!
You may then open the rotor e.g. in a safety work
bench and decontaminate it.
Audible alarm
Accompanying all error messages, a warning signal is
given out which only is silenced upon actuating any
key.
As an option, you can also have the end of a run sig-
naled acoustically. For activating or deactivating this
option please press the
key in the actual value
mode. After about 1 sec you get into the selection
mode. The speed display shows “beep” and the time
display shows “on” or “off”. As long as the key is
pressed, the selection mode is active.
By actuating the upward-key
in the time display
the signal function can be switched on or off.
When the message “rotor” flashes, press-
ing the start key once is sufficient to turn
off the warning signal and to accelerate the
rotor to the maximum speed displayed by
the instrument.