Rotors and accessories
37
BIOshield
®
-Rotor 75006435
Highplate
®
-Rotor 75006444
Closing round buckets aerosol-tightly
After greasing the seal, turn the lid until it fits slightly
to the bucket.
To achieve an uniform pre-stressing, turn the lid
clockwise by 1 ½ grip areas ( approx. 15°). Use the
setting marks on the bucket as an orientation.
If the windshield-rotors are filled with samples outside
the centrifuge (e.g. in a safety work bench), it is es-
sential that during the insertion into the centrifuge the
rotor is kept in a vertical position.
Grip areas
Bucket mark
The aerosol-tight bio-containment of
windshield-rotors is only secured in a
horizontal position!