Maintenance and care
64
Never exceed the maximum per-
missible autoclaving time and tem-
perature.
Aluminum rotors have to be treated
with corrosion protective oil each
time they have been autoclaved.
Should the rotor show signs of
wear or corrosion, you must stop
using it!
Decontamination
For general radioactive decontamination, use a solu-
tion of equal parts of 70
% ethanol, 10
% SDS
and water. Follow this with ethanol rinses, then de-
ionized water rinses, and dry with a soft absorbent
cloth. Dis- pose of all washing solutions in appropriate
radioactive waste containers!
Autoclaving
Check whether autoclaving
is permitted!
(see label on the rotor body and the rotor lid)
For the autoclavable parts the permissible autoclaving
cycle is 20 min at 121
°C.
Remove the rotor lid together with the centrifuge
tubes and the adapter. The rotor has to be cleaned
and rinsed with distilled water before being auto-
claved. Buckets and tubes must not be closed during
autoclaving.
Place plastic rotors on an even surface to avoid de-
formation.
Chemical additives to the steam are not
permitted.
Corrosion protective oil 70009824 is delivered with
the centrifuge.