Before the use
15
The mains switch must not be used to
interrupt the centrifuge run (except for
emergencies). For this, the STOP key
is provided!
Connection conditions
Before the centrifuge is connected with the mains net-
work, the following must be ensured:
1. A water connection is available for a ¾" screwed
connection:
Min. flow rate of the water 10 l
itres
per
minute Water pressure from 2.5 bar to 6.0
bar Water temperature 14
°C to 18
°C
2. Water runoff (pressure-free) is available
3. The voltage information on the type designation
plate is identical to the available mains voltage
4. The mains connection at a plug socket with protec-
tive grounding takes place.
The Cryofuge GMP can be connected
to a closed-circuit in-house water
circulation system. The connection to
a water-circulation-cooler (e.g. HWK
4600/A, Order No. 76007473) is also
possible.
Commissioning
Connect plug with plug socket.
Ensure water supply and return flow.
Activate mains switches in front to the right.
All strainer segment displays and diodes light up after
that for short time for the purpose of verification
(system check).
The hoses for the water connection
may be exchanged only by technical
personnel.
Shortening of the hoses is not
permissible!
Содержание Cryofuge 5500i GMP
Страница 16: ...Description of the Cryofuge 12 Notes ...
Страница 20: ...Before the use 16 Notes ...
Страница 70: ...Maintenance and care 66 ...
Страница 99: ...Appendix 95 ...
Страница 107: ...Appendix 103 ...