CO
2
Incubator BB 150 Operating Instructions
59
10.
Maintenance
10.9
Replacing the door seal
The door seal (magnetic seal) of the outer door is located in the
retaining slot. No tools are required to replace the seal.
1.
Fig. 27:
Pull magnetic seal [3] out of the guide slot [1].
2. Position new seal at a corner [2] and press seal retaining
rail [4] into slot.
3. Make sure that the retaining rail taper is positioned correctly
in the slot [1] and that the seal is flush with the door frame.
Fig. 27
Door seal replacement