4.4
SWITCHING THE UNIT ON
The following actions are initiated when the unit is switched on:
The green toggle switch on the main power control indicates that power to the
unit has been turned on.
Both displays perform an 8-digit test, and all LEDs come on.
When high/low humidity, O
2
, and/or flask rotation equipment are configured
their LEDs also light up during the 8-digit check.
Options that are not configured do no light.
The version information of the various software versions is then displayed:
P 1 (operator and display PCB):
e.g., 010 = Version 3
P 2 (sensor PCB):
e.g., 026 = Version 27
P 3 (main board):
e.g., 204 = Version 300
The unit then displays the current actual values.
The selected unit version is shown after the three version numbers during
initialization. If the unit version comprises more than three figures, the full
number appears in two areas on the display.
Example:
Pr (unit version) = 4608
→
First, “004” is displayed,
followed by “608” on the display.
Please provide the program version when making inquiries, particularly with regard to
avoidable software problems.