CO
2
Incubator BB 150 Operating Instructions
26
Fig. 14
Gas pressure hose
installation,
CO
2
connection
5.
Start-up
5.4
Gas connection
NOTE – Gas quality
CO
2
must be of the following qualities:
• Purity 99.5 % min,
• medical grade.
CAUTION – Overpressure!
The operating pressure of the gas applied to
the device must not exceed 1 bar. If the gas is
supplied at a higher pressure, the valves inte-
gral to the device may not close correctly and
the gas supply control may be impaired.
Set the gas supply to a range between 0.8 bar
min. and 1.0 bar max. and make sure that this
pressure setting cannot be changed!
NOTE – Pressure compensation opening
To ensure permanent pressure compensation,
the pressure compensation opening must not
be connected to an exhaust air system. The
pipe of the pressure compensation opening
must not be extended or redirected.
NOTE – Access port
If the access port is not used, it must be cap-
ped during operation.
CO
2
connection
Fig. 14:
The gas supply from the gas supply system to the
device is achieved using the supplied flexible gas pressure
hoses:
1. Connect gas pressure hose [1] to the sleeve of the gas sup-
ply system.
2. Remove protective cap [3] from sterile filter.
3. Slide hose clamp [2] onto gas pressure hose and connect
hose to the sleeve of the sterile filter.
4. Secure gas pressure hose to the sleeve of the sterile filter
using the hose clamp.