Name
Date
Designation
Prepared
JF 17.07.13
Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50079823_09_CN
50 079 823
2/ 2
100
Temperature below set value Actual value < set value - 1.0 °C
101
Temperature above set value Actual value > set value + 1.0 °C
(Sample protection function active.)
104
Temperature probe PT1000
or digital temperature sensor
LM 75 defective.
Probe break or sensor short-circuit
200
CO
2
below set value
Actual value < set value - 1.0 % CO
2
201
CO
2
above set value
Actual value > set value + 1.0 % CO
2
204
CO
2
measuring cell defective Sensor break or short-circuit, or infrared
measuring cell defective.