Incubator BB 150 Operating Instructions
50
9.
Cleaning and disinfection
Final disinfection:
1 Install shelf system and accessories.
2. Again, spray disinfectant onto the surfaces of the work space and of the ac-
cessories or wipe the surfaces clean using disinfectant.
3. Allow disinfectant to react as speciefied by manufacturer.
9.3
ContraCon decontamination routine
CAUTION! – Hot surface!
The surfaces of the work space, particularly the
glass door armatures and the interior sheet of
the outer door, are heated during the
decontamination routine.
During the routine run or immediately after
completion of the run, always wear safety
gloves when touching these surfaces; observe
the warning indicator at the operating panel!
The entire program run of the decontamination routine takes approx 25 hours.
1. After the cleaning, reinstall the shelf system components into the work space.
2. Fill the water tray with 300 ml of processed water.
3. Turn the device on using the power switch.
4. Activate decontamination routine (see table in Section 9.4).
5. After the decontamination routine has been completed, remove the remaining
water using a sterile cloth.
6. Turn the device off or restart the device operation using auto-start
(see Section 7.2).
NOTE – Duration of the auto-start routine
When the device is cold and when the
ambient temperature is low, the auto-start
routine may take up to 10 hours.
ContraCon decontamination routine procedure:
Fig. 22:
The routine is divided into four phases. Each individual phase or several
phases can be cancelled (i.e. skipped).
If the operating step ”Cancel ContraCon routine” is executed, the routine moves
to the next program phase. To cancel the routine completely, the operating step
must be executed repeatedly until the remaining run time display shows the
value 0.
When the glass door is opened, this operating step cancels the routine completely.
The remaining run time of the ContraCon decontamination routine designates the
period between the start or the current routine time state and the cooling down to
the preset temperature nominal value (± 2° C).