CO
2
Incubator BB 150 Operating Instructions
15
3.
Installation of the device
3.5
Stacking
Fig. 3:
Two BB 150 devices can be stacked on top of each other.
The stacking elements [1] and [2] secure the devices on top
of each other.
If the devices are placed onto mobile racks, ensure that the
rollers [4] are secured by an arresting device during operati-
on. For reasons of stability, the rollers should be oriented to
the front.
When stacking the incubators, the stacked incubator function
must be activated on the upper unit. This function adjusts the
parameters of the uppper unit to compensate any heat trans-
mission between the two units.
NOTE – Transporting stacked devices
The stacking elements are anot connecting
elements. Therefore, the transport of stacked
devices on sloped surfaces is not allowed.
3.6
Modifications
The outer door and the glass door can be equipped with left
or right side fasteners. The door fastening can also be reversed
later.
NOTE – Modifications
Retrofitts and modifications must only be per-
formed by the Technical Service of Thermo
Fisher Scientific Products.
Fig. 3
Stacking