Name
Date
Designation
Prepared
JF 05.07.13
Service Manual BB 150
Index
Name
Date
Document No.
Page
File
50078923_04_CN
50 079 823
30/ 30
4.7.8 Faults without Error Messages
Error
code
Faulty function
Inspections and tests
Possible corrective actions
- Heavy
condensation in
the sample
chamber
Check the sample
chamber heater
actuators (heaters
E2-E8).
Check the unit
version.
Replace the main board.
Return to factory for repairs.
Set correct set of parameters
(FL 21)
- Heavy
condensation on
the glass
Check the door
heater actuator
(heater E1).
Check
temperature
switch N2 in the
exterior door.
For software version
<200: check door
factor (FL 8)
Check set of
parameters (since
software version
200)
Replace the main board.
Replace the entire exterior
door.
Set correct door factor
Set correct set of parameters
(since software version 200,
FL 21)